This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
removing_air_bubbles_lines [2020/09/25 00:08] richard |
removing_air_bubbles_lines [2021/08/16 22:22] (current) |
||
---|---|---|---|
Line 13: | Line 13: | ||
// | // | ||
// | // | ||
- | {{youtube> | + | {{youtube> |
Line 42: | Line 42: | ||
---- | ---- | ||
+ | |||
+ | ==== Tubing perfusion when 4-way valve in output position ==== | ||
< | < | ||
Line 47: | Line 49: | ||
</ | </ | ||
+ | This will clear the lines from the PCV to the pipettes but you still need to get the air bubbles out between the syringes and valves. To do this, now you need to: | ||
+ | |||
+ | - Position the corresponding valve into the output position. | ||
+ | - Draw some fluid into the syringe connected to the manual valve that draws fluid from the PCV. | ||
+ | - Disconnect the tubing from one of the multichannel syringes and put that end in a bottle and disconnect the corresponding tubing to pipette and attach the fluid filled syringe. | ||
+ | - Perfuse until all air bubbles are out and attach to multichannel syringe pump first, put valve into bypass position, then attach the other end of tubing to the corresponding pipette. | ||
+ | |||
+ | ---- | ||
+ | < | ||
+ | {{ :: | ||
+ | </ |