Differences
This shows you the differences between two versions of the page.
Both sides previous revision
Previous revision
Next revision
|
Previous revision
|
removing_air_bubbles_lines [2020/09/25 00:13] richard |
removing_air_bubbles_lines [2021/08/16 22:22] (current) |
// | // |
// | // |
{{youtube>GeB6QfONg6w}} | {{youtube>GeB6QfONg6w?&rel=0 }} |
| |
| |
This will clear the lines from the PCV to the pipettes but you still need to get the air bubbles out between the syringes and valves. To do this, now you need to: | This will clear the lines from the PCV to the pipettes but you still need to get the air bubbles out between the syringes and valves. To do this, now you need to: |
| |
* Position the corresponding valve into the output position. | - Position the corresponding valve into the output position. |
* Draw some fluid into the syringe connected to the manual valve that draws fluid from the PCV. | - Draw some fluid into the syringe connected to the manual valve that draws fluid from the PCV. |
* Disconnect the tubing from one of the multichannel syringes and put that end in a bottle and disconnect the corresponding tubing to pipette and attach the fluid filled syringe. | - Disconnect the tubing from one of the multichannel syringes and put that end in a bottle and disconnect the corresponding tubing to pipette and attach the fluid filled syringe. |
* Perfuse until all air bubbles are out and attach to multichannel syringe pump first, put valve into bypass position, then attach the other end of tubing to the corresponding pipette. | - Perfuse until all air bubbles are out and attach to multichannel syringe pump first, put valve into bypass position, then attach the other end of tubing to the corresponding pipette. |
| |
| ---- |
| <figure> |
| {{ ::labautobox_attach_filled_syringe_to_one_syringepump_channel_v2.png?400|To remove the bubbles coming out of the multichannel syringe pump to the pipettes. You need to fill up a syringe with fluid, disconnect the tubing from the pipettes, disconnect the corresponding tubing that is connected to one of the syringes in the multichannel syringe pump, connect that end to a fluid reservoir (like a bottle) then push bubbles out.}}<caption>**To remove the bubbles coming out of the multichannel syringe pump to the pipettes. You need to fill up a syringe with fluid, disconnect the tubing from the pipettes, disconnect the corresponding tubing that is connected to one of the syringes in the multichannel syringe pump, connect that end to a fluid reservoir (like a bottle) then push bubbles out.**</caption> |
| </figure> |