====== Putting together a pipette flushing macro ====== Flushing the pipette tips is a good way to assess that they air bubbles are out of the lines everything is working well. Here are steps on how to build a flushing macro using the "[[software_overview|Build Macro]]" page (introduced here), which will be saved as a custom macro. Generally, the steps involved are as followed: - Move 4-way valves to bypass - Move pipettes to waste station - Move 4-way valves to input - Load multichannel syringes with 500 microliters - Turn on waste pump - Move 4-way valves to output - Dispense 500 microliters - Dry pipette tips - Turn off waste pump ---- Here is a video that explains how to compile the macro. // // {{youtube>w-urOVQg5V4}} ---- Here is the macro code. // // //valve-1_1_1_1_1_1_1_1-valvebypass valve-1_1_1_1_1_1_1_1-bypass //motion waste station zh 0 row 1 F3000 G1Z0F3000 G1X235.5Y24F3000 G1Z12F3000 //valve-1_1_1_1_1_1_1_1-valveinput valve-1_1_1_1_1_1_1_1-input //syringe 500 speed 40 acc 20 sg1e500s40a20_17 G1Z12F3000_2 //waste 2 on wasteon //valve-1_1_1_1_1_1_1_1-valveoutput valve-1_1_1_1_1_1_1_1-output //syringe 0 speed 40 acc 20 sg1e0s40a20_17 //valve-1_1_1_1_1_1_1_1-valvebypass valve-1_1_1_1_1_1_1_1-bypass //just dry function touchdry Z9.5F3000ZT0T0.1 //waste 2 off wasteoff ---- Here is video showing the process // // {{youtube>826EpRLjeEA?&rel=0|Pipette flushing example}}